Care and maintenance

4. Handling Errors and Problems

4.1

Error Messages

Error Display
EE displayed in the
systolic pressure
area.
E displayed in the
systolic pressure
area.
EE displayed.
Note: If you change your posture, or move, during measurement, the posture
symbol (
measurement.
Cause
Movement during
measurement
Is the arm placed through
the arm cuff correctly?
Are rolled up sleeves
squeezing the arm?
Is air leaking from the arm
cuff?
Movement during
measurement
Arm cuff has been inflated
above 299 mmHg.
) or movement symbol (
4. Handling Errors and Problems
Remedy
Check that the heartbeat symbol is
lit, then keep still and repeat
measurement.
(Refer to Chapter 3.3)
If the EE appears again, use again,
refer to Chapter 3.4.
Place your arm correctly.
(Refer to Chapter 3.2)
Remove clothing from your upper
arm and place your arm through the
cuff. (Refer to Chapter 3.2)
Contact your OMRON distributor or
Customer Services as mentioned
on the package.
Check that the heartbeat symbol is
lit, then keep still and repeat
measurement.
Do not inflate the arm cuff above
299 mmHg.
(Refer to Chapter 3.4)
) may be displayed during
EN
39