In order to restrict access of certain settings to the medical team, a hidden key has been provided on
the device. It is located under the letters "Go" in the "GoodKnight" name. This is detectable to the
touch as a hollow.
The parameter values may be changed by using the on/off and ramp keys as shown in the figure
"View of the control buttons" below:
Hidden key (allows to move
back through the menu)
Decreases parameter
values or enables / disables
some parameters.
Information Access
button
NOTE
The layout of the buttons is the same on the GoodKnight 425 ST as on the GoodKnight 425, only
the device's name is different.
To access all the settings, press the hidden key for one second and then simultaneously the
information access key
NOTES
· The symbol reminds you that the parameter value is adjustable.
· All the values displayed in the following screens are given as a guide only.
Press the hidden
The screen displays:
key for one
second and
key .
With each additional press of the information access key, the following settings will be available.
EN – 6
View of the control buttons
.
Back-up rate of the device.
F 10
This function is accessible only on the GoodKnight 425
ST device.
Use the on/off or ramp key to increase or decrease the
Back-up rate value. This value is adjustable between 4
and 25 bpm.
If 0 is displayed, it means that there is no Back-up rate.
NOTE
GoodKnight 425 / GoodKnight 425 ST
Increases parameter
values or enables /
disables some parameters.