C e F A R t e n s
C a u t ion !
- It is recommended to use the transport bag that comes with the unit, for transport
of the device, and to use a proper transport box to ship it.
- Always use the AC adaptor (power supply) provided by the manufacturer to recharge
the unit. Never use any other AC Adapter. This could damage the device and
generate electrical hazards.
- Only use electrodes supplied by the manufacturer. Other electrodes may have
electrical properties that are unsuitable for or may damage the CEFAR TENS.
- Do not place the electrodes in water.
- Do not apply solvents of any kind to the electrodes.
- Instructions for electrodes. See the usage and storage instructions displayed on the
bag of electrodes.
- Hot casing or battery. Under extreme use conditions, some parts of the casing
might reach up to 118 °F (48 °C). Use caution when manipulating the rechargeable
battery right after device use or when holding the device. There is no particular
health risk associated with this temperature besides your comfort
- size of electrodes. Do not use electrodes with an active area of less than 10 cm²
due to the risk of associated burning. Proceed systematically with caution when the
density of the current is over 2 mA/cm².
- If the device is not used for an extended period, we recommend that you charge your
device every 5 months. You should store the device in a dry and cool environment.
no t e
- For best results, wash and clean the skin of any oil and dry it before attaching the electrodes.
- Never use a set of adhesive electrodes for more than 40 sessions as the quality of the contact
between the electrode and the skin, which is essential for the patient's comfort and the
effectiveness of the stimulation, gradually deteriorates.
- For information on use and storage please consult the instructions found on the electrodes
packaging.
no t e
Biocompatibility
Those parts of the CEFAR TENS unit that come into contact with the patient when used as
intended, are designed to fulfil the biocompatibility requirements of the applicable standards.
e n