see figure 14:
.­ Technical­data­of­the­measuring­accessories
­
4­mm­Safety­measuring­cable­ATL­2
-
S tandard: EN 61010-031,
-
M aximum rated voltage to earth (­ ) and measuring category: 1000 V
CAT III, 600 V CAT IV,
-
M aximum rated current: 10 A,
-
P rotective class II (­  ), continuous double or reinforced insulation,
-
C ontamination class: 2,
-
L ength: 1.4 m, AWG 18,
-
E nvironmental conditions:
Maximum barometric elevation for making measurements: 2000 m,
Temperatures: 0 °C to + 50 °C, humidity 50 % to 80 %
-
O nly use the measuring cables if in perfect condition and according to this
manual, since the protection provided could otherwise be impaired.
-
T hrow the measuring cable out if the insulation is damaged or if there is a
break in the cable/ plug.
-
D o not touch the bare contact tips of the measuring cable. Only grab the
area appropriate for hands!
-
I nsert the angled terminals in the testing or measuring device.
2.­ Environmental­notice
At the end of the product's useful life, please dispose of it at
appropriate collection points provided in your country.
08/ 2007
Assembly of the BENNING MM 8/ 9/ 10
­ B ENNING­MM­8/­9/­10­
20