(­BENNING MM 9/ 10). Place the battery/batteries into the battery compart-
ment provided for this purpose.
-
P lace the battery compartment cover onto the bottom part and tighten the
screws.
-
I nsert the BENNING MM 8/ 9/ 10 into the rubber protection frame .
see figure 11:
Make­ your­ contribution­ to­ environmental­ protection!­ Do­ not­
dispose­ of­ discharged­ batteries­ in the household garbage.
Instead,­ take­ them­ to­ a­ collecting­ point­ for­ discharged­
batteries­ and­ special­ waste­ material.­ Please­ inform­ yourself
in­your­community.
9.4­ Fuse­replacement­(­BENNING MM 9/ 10)
Before­opening­the­BENNING­MM­9/­0­make­quite­sure­that it
is­voltage­free!­Electrical danger!
The BENNING MM 9/ 10 is protected against overload conditions by means of
a fitted fuse (­G melt insert) 10 A (­see Figure 12)
Proceed as follows to replace the fuse:
-
Disconnect the safety measuring cables from the measuring circuit.
-
Disconnect the safety measuring cables from the BENNING MM 9/10/.
-
Turn the rotary switch  to the switch setting "OFF".
-
Remove the rubber protection frame  from the BENNING MM 9/ 10.
- L ay the BENNING MM 9/ 10 face down and release the four slot screws of
the housing bottom.
Do­ not­ undo­ any­ screws­ on­ the­ printed­ circuit­ of­ the­ ­
­ B ENNING­MM­9/­0!
-
L ift up the housing bottom in the lower section and remove the same at the
top from the front section.
- L ift one end of the defective fuse from the fuse holder.
-
F ully push the defective fuse out of the fuse holder.
- I nsert a new fuse which has the same current rating, disconnecting charac-
teristic and dimensions.
-
M ake sure that the new fuse is seated centred in the holder.
-
O rder the battery supply lines such that they are not crushed between the
housing parts.
-
E ngage the front end on the housing bottom and fit the three screws.
-
Insert the BENNING MM 9/ 10 into the rubber protection frame .
see figure 12:
9.5­ Calibration
To maintain the specified precision of the measurement results, the instrument
must be recalibrated at regular intervals by our factory service. We recom-
mend a recalibration interval of one year. Send the appliance to the following
address:
Benning Elektrotechnik & Elektronik GmbH & CO. KG
Service Centre
Robert-Bosch-Str. 20
D - 46397 Bocholt
9.6­ Spare­parts
Fuse F 10 A, 500 V, 1.5 kA, D = 6.3 mm, L = 32 mm
0.­ Application­of­rubber­protection­frame
-
Y ou can store the safety measuring lines by winding the safety measuring
lines around the rubber protection frame  and engaging the tips of the
safety measuring lines with protection against the rubber protection frame
 (­see figure 13).
-
Y ou can engage a safety measuring line to the rubber protection frame
, that the measuring tip protrudes freely, in order to be able to apply the
measuring tip - jointly with the BENNING MM 8/ 9/ 10 to a measuring point.
-
T he rear support on the rubber protection frame  provides for an inclined
setup of the BENNING MM 8/ 9/ 10 (­facilitates reading) or attachment (­see
Figure 14).
T he rubber protection frame  is fitted with a lug which can be freely used
for an attachment option.
see figure 13:
08/ 2007
Battery change
Electrical­danger!
Fuse replacement
Winding up of the safety measurement line
­ B ENNING­MM­8/­9/­10­
in­ the­ household­ garbage.­
yourself­
that­it­
P.no. 749726
9