(BENNING MM 9/ 10). Place the battery/batteries into the battery compart-
ment provided for this purpose.
-
P lace the battery compartment cover onto the bottom part and tighten the
screws.
-
I nsert the BENNING MM 8/ 9/ 10 into the rubber protection frame .
see figure 11:
Make your contribution to environmental protection! Do not
dispose of discharged batteries in the household garbage.
Instead, take them to a collecting point for discharged
batteries and special waste material. Please inform yourself
inyourcommunity.
9.4 Fusereplacement(BENNING MM 9/ 10)
BeforeopeningtheBENNINGMM9/0makequitesurethat it
isvoltagefree!Electrical danger!
The BENNING MM 9/ 10 is protected against overload conditions by means of
a fitted fuse (G melt insert) 10 A (see Figure 12)
Proceed as follows to replace the fuse:
-
Disconnect the safety measuring cables from the measuring circuit.
-
Disconnect the safety measuring cables from the BENNING MM 9/10/.
-
Turn the rotary switch to the switch setting "OFF".
-
Remove the rubber protection frame from the BENNING MM 9/ 10.
- L ay the BENNING MM 9/ 10 face down and release the four slot screws of
the housing bottom.
Do not undo any screws on the printed circuit of the
B ENNINGMM9/0!
-
L ift up the housing bottom in the lower section and remove the same at the
top from the front section.
- L ift one end of the defective fuse from the fuse holder.
-
F ully push the defective fuse out of the fuse holder.
- I nsert a new fuse which has the same current rating, disconnecting charac-
teristic and dimensions.
-
M ake sure that the new fuse is seated centred in the holder.
-
O rder the battery supply lines such that they are not crushed between the
housing parts.
-
E ngage the front end on the housing bottom and fit the three screws.
-
Insert the BENNING MM 9/ 10 into the rubber protection frame .
see figure 12:
9.5 Calibration
To maintain the specified precision of the measurement results, the instrument
must be recalibrated at regular intervals by our factory service. We recom-
mend a recalibration interval of one year. Send the appliance to the following
address:
Benning Elektrotechnik & Elektronik GmbH & CO. KG
Service Centre
Robert-Bosch-Str. 20
D - 46397 Bocholt
9.6 Spareparts
Fuse F 10 A, 500 V, 1.5 kA, D = 6.3 mm, L = 32 mm
0. Applicationofrubberprotectionframe
-
Y ou can store the safety measuring lines by winding the safety measuring
lines around the rubber protection frame and engaging the tips of the
safety measuring lines with protection against the rubber protection frame
(see figure 13).
-
Y ou can engage a safety measuring line to the rubber protection frame
, that the measuring tip protrudes freely, in order to be able to apply the
measuring tip - jointly with the BENNING MM 8/ 9/ 10 to a measuring point.
-
T he rear support on the rubber protection frame provides for an inclined
setup of the BENNING MM 8/ 9/ 10 (facilitates reading) or attachment (see
Figure 14).
T he rubber protection frame is fitted with a lug which can be freely used
for an attachment option.
see figure 13:
08/ 2007
Battery change
Electricaldanger!
Fuse replacement
Winding up of the safety measurement line
B ENNINGMM8/9/10
in the household garbage.
yourself
thatit
P.no. 749726
9