DC AND AC CURRENT MEASUREMENT (See Fig. 2)
❶ Connect red test lead to the mA Input for current measurements up to 200mA
or to the 20A input for current measurements to 20A (15 to 20A for 30 seconds).
Connect black test lead to COM Input Connector. ❷ Set the Function/Range
switch the desired DC or AC current range corresponding to the selected input
jack. If in doubt choose to the highest setting first. ❸ Open circuit in which
current is to be measured. Securely connect test leads in series with the load. ❹
Turn on power to circuit being tested. Read current value on Digital Display.
Note: Each current input has a protective fuse installed. If fuse does blow, replace
with same.
Note for AC Measurements: See AC/DC Voltage measurement.