4.2
General weighing information
S The patient to be weighed must sit without support in the load suspension
device (e.g. sling). Please note that the permissible total load of the scale
must not be exceeded.
S Neither the patient nor the suspension device must be touched during
weighing. The cradle should not swing back and forth.
S The weight value should only be read once the scale is at rest. This is
indicated when the kg weight unit shows continuously on the display.
S If there is too much patient movement to allow the display to come to a
standstill, modifying the filter setting as per section 7.2 can remedy the
situation.
4.3
Weighing when switching the scale on without a
patient
Once the segment test is completed as per section 4.1, the display is set to
0.0 kg and the scale is ready to use.
Before the patient can sit in the scale, the weight that is not to be included in
the measurement (e.g. a towel) must be placed on the scale and tare key G/N
>0/T< must be pressed.
1 x
SLS
HBM Short Reference Guide Lifter Scale A1221-5.2 Issue July 31, 2014
"
Press the G/N key once to tare
You can now weigh
"
Press the G/N key once again to toggle between the
scale zero setting and the gross weight.
"
Only then should the patient sit in the sling. The
weight can now be read on the display.
Weighing
9