Note: If the sensor cable is not long enough to reach the controller, an interconnect cable (model 370=506=025)
and pre-amplifier unit (model 08350=A=8000) should be used to extend the distance.
Terminal
8-pin (J5)
1
2
3
4
5
6
7
8
2-pin (J4)
1
2
Terminal
8-pin (J5)
2-pin (J4)
In Step 3 of the following sequence, set the switch as follows:
Switch Setting
Setting 2
Setting 3
Setting 6
Setting 7
6 English
Table 1 pH and ORP sensor wiring
Description
Reference
Ground solution
Jumper 1-2 on J5
—
—
—
Temp -
Temp +
—
Active
—
Table 2 Antimony electrode sensor wiring
1
2
3
4
5
6
7
8
1
2
8350/8351
8362
Black
Green
Jumper 1-2 on J5
—
—
—
White
White
Red
Brown
—
Transparent
Transparent
—
Description
Reference
Ground solution
—
—
—
Temp -
Temp +
—
Active
—
Description
pH combination sensor
ORP combination sensor
Antimony electrode sensor
User defined
8417
White
Jumper 1-2 on J5
—
—
—
—
—
—
White
Brown
—
—
Green
—
—
8346/8347
Black
Jumper 1-2 on J5
—
—
—
White
Violet
—
Red
—