P3.6 Ion Sensor ID
Enter this menu to assign an identifier name to the sensor currently in use.
To activate this option, the external keyboard must be connected.
The sensor ID will appear in the "Complete" display screen (P6.7) and in Complete or GLP paper print
(P5.3)
P3.7 Restore factory default.
If the instrument does not work properly or incorrect calibrations have been performed press Yes with
key to return all Ion parameters to the factory default setting.
IMPORTANT: The factory reset of the parameters does not delete the stored data.

• Calibration with Ion-Selective Electrodes

Example for 2 points calibration 0.01 and 0.1 mg/L
Enter in the Ion setup and select in program 3.1 the measuring unit mg/L and in the 3.2 the most
diluted standard: 0.010
The device automatically multiplies the lower standard for a factor by a 10 to locate the other
points in the calibration curve.
Connect the appropriate ISE electrode to the connector for pH/mV/ORP (yellow/blue).
Important: If the ISE electrode is not combined, it is necessary to connect the specific reference electrode.
Refer to the ISE electrode manual for any reference electrode filling electrolytes and for any ionic strength
adjusters (ISA).
Press
key to get back to measurement mode and shift to MEAS ION with
Press and hold for 3 sec "
Rinse the electrode with distilled water
Press
and immerse the electrode in the most diluted standard
(Low Standard P3.2) as indicated by becker
When
icon appears (or when time runs out if you chose
"Time" as the criterion for stability) press
first calibration point.
Extract the electrode, rinse with distilled water
Press
to continue and immerse the electrode in the next standard (Low Standard X 10) as indicated
0.100
by becker
.
When
icon appears press
When the second point is finished, the user can exit calibration by pressing
to continue with the remaining points.
When the calibration is completed, the calibration report is displayed with CALIBRATION DATE and
HOUR / TEMPERATURE / MEASURING UNIT / ID SENSOR / SLOPE % for every range.
UM Serie 880 EN rev.2 30.07.2020
CAL" key to enter into the calibration mode.
and gently dab with absorbent paper.
0.010
.
key to confirm the
and gently dab with absorbent paper.
key to confirm the second calibration point.
21
key.
key; instead press