Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 6)
Recommended separation distances between portable and mobile RF
communications equipment and the electric breastpump Mini Electric
The electric breastpump Mini Electric is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the electric breast-
pump Mini Electric can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the
electric breastpump Mini Electric as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum
output power of
transmitter W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
Note 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Separation distance according to frequency of
transmitter M
150 kHz to 80 MHz
d = 1,2√P
0,12
0,37
1,2
3,7
12
80 MHz to
800 MHz
d = 1,2√P
0,12
0,23
0,37
0,74
1,2
2,3
3,7
7,4
12
23
800 MHz to 2,5 GHz
d = 2,3√P