Install The Lamination Dummy - Ottobock Lamination Anchors Instrucciones De Uso

Ocultar thumbs Ver también para Lamination Anchors:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 37

5.3 Install the lamination dummy

INFORMATION
► Use the 4X46 lamination dummy when the lamination anchor is
to be used with a screwed insert.
► Use the 4X46=ST lamination dummy when the next prosthetic
component is to be screwed directly into the lamination anchor
(e. g. 4R57=ST). The lamination dummy is somewhat higher
proximally so that the prosthetic component does not contact the
laminate when fully screwed in.
Lamination anchor without threaded connector
► Set the 4X3 lamination dummy on the pyramid adapter of the lam­
ination anchor (see fig. 12).
Lamination anchor with threaded connector
1) Tighten the cap screw of the threaded clamping to a point where
the screwed insert can still be screwed in and out.
2) Unscrew the screwed insert from the lamination anchor.
3) Screw in the 4X46* lamination dummy (see fig. 13).
5.4 Assembling the distal prosthetic components
Installing the screwed insert
The threaded connector is used for installing a connecting component
with an external thread (e.g. pyramid adapter or prosthetic knee joint).
Required materials: 710D4 torque wrench, 636K13 Loctite®
>
1) Screw in the screwed insert (e.g. pyramid adapter) completely
into the thread of the lamination anchor.
2) The
screwed
insert
um of 1/4 turn out of the thread in order to align the insert.
3) If a titanium screw is on hand: After trial fitting, replace the
steel cap screw with the titanium cap screw.
4) Secure the cap screw of the threaded clamping using Loctite®
and tighten using the torque wrench (10 Nm).
18
may
be
unscrewed
by
Connecting the pyramid adapter and pyramid receiver
The pyramid adapter is fixed with the set screws of the pyramid receiv­
er.
Required materials: torque wrench (e.g. 710D20), 636K13 Loc­
>
tite 241
1) Fitting:
Screw in the set screws.
Use the torque wrench to tighten the set screws (10 Nm).
2) Definitive assembly:
Secure the set screws with Loctite.
Screw in the set screws.
Pre-tighten the set screws with the torque wrench (10 Nm) and
then tighten them (15 Nm).
3) Replace any set screws that are protruding or are recessed too
much with appropriate set screws (see selection chart).
Reference number
506G3=M8X12-V
506G3=M8X14
506G3=M8X16
Alignment
The set screws in the pyramid receiver can be used to make static
adjustments during alignment, trial fittings and after the prosthesis is
finished.
Replacement and disassembly
maxim­
The set position of the prosthetic component can be maintained dur­
a
ing replacement or disassembly. In order to do this, unscrew the two
set screws that are screwed in the furthest and located next to each
other.
6 Cleaning
1) Clean the product with a damp, soft cloth.
2) Dry the product with a soft cloth.
Selection table for set screws
Length (mm)
12
14
16
Tabla de contenido
loading

Tabla de contenido