To ensure compliance with the EMC requirements as specified in IEC 60601-1-2, it is essential to use
only the following accessories:
EUT Support Equipment
ITEM
DC Power Supply
Firewire® PCIExpress Card
EUT Support Cables
Description
DC Power Supply
Firewire® PCIExpress
Card
This VisualEyes™ is suitable in hospital environments except for near active HF surgical equipment and
RF shielded rooms of systems for magnetic resonance imaging, where the intensity of electromagnetic
disturbance is high.
Use of accessories, transducers and cables, other than those specified or provided by the manufacturer
of this equipment, could result in increased electromagnetic emissions or decreased electromagnetic
immunity of this equipment and result in improper operation. The list of accessories, transducers and
cables can be found in the EMC appendix of this instruction.
Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of this VisualEyes™, including
cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could
result.