R&S
-Z3
Performance Test of Return Loss at 320 MHz to 340 MHz Range
4.4
Performance Test of Return Loss at 320 MHz to
340 MHz Range
To perform the test
1. Adjust the network analyzer settings:
● Source Power: 0 dBm
● Ref. Level: 0 dB
● Ref. Attenuator: 0 dB
● Center frequency: f
● Span: 150 MHz
2. Insert the short antenna rods on the side of the antenna head using the snap-
in connectors.
3. On the network analyzer, set markers at:
● 320 MHz
● 330 MHz
● 340 MHz
4. Measure the S11 parameter. It should be ≤ -12 dB (see
5. On the network analyzer, change the frequency range to:
● f
= 320 MHz
start
● f
= 340 MHz
stop
6. Set the marker to maximum mode.
7. Measure the S11 parameter. It should be ≤ -12 dB (see
Figure 4-2: Typical return loss at 320 MHz to 330 MHz range
Manual 5202.9444.02 ─ 01
= 330 MHz
center
Functional Test
Figure
4-2).
Figure
4-2).