Intensity:
Tolerances:
Display:
Calibration:
Blood pressure (NIBP)
Cuffs:
Measurement range:
Measurement range for
neonates:
Measurement precision:
Leakage:
Overpressure limit:
Overpressure limit (2nd level):
Time per measurement:

Scope of delivery / accessories

Scope of delivery
ri-vital
®
Mains adapter
Fingerclip sensor
Blood pressure connector tube
Blood pressure cuff, size depending on order number
Operating manual
Accessories
Finger sensor adults
Finger sensor children
Sleeve sensor neonates
Velcro cuff neonates
Velcro cuff infants
32
3.8 mW maximum
SpO2
+/- 2 % for 70 - 100 %
(below this unspecified)
Pulse
+/- 1% or +/- 1 position
Oxygen saturation
(Updating of the readings once per second), pulse as well
as a bar graph for pulsation (display is not proportional to
the strength of the pulse wave.)
Functional saturation
All cuffs from the Riester portfolio
Systole
25 - 280 mmHg
Diastole
10 - 200 mmHg
MAP
15 - 260 mmHg
Systole
20 - 155 mmHg
Diastole
5 - 110 mmHg
MAP
10 - 130 mmHg
Mean value difference < +/-5mmHg
Standard deviation < 8mmHg
< 3 mmHg / minute
300 mmHg in "Adult" mode
150 mmHg in "Neo" mode (neonates)
330 mmHg in "Adult" mode
165 mmHg in "Neo" mode (neonates)
approx. 25s on average
"Adult" mode max. 170 s
"Neo" mode max. 80s
(article number 12661) patient weight > 20 Kg
(article number 12662) patient weight > 20 Kg
(article number 12663) patient weight > 1 Kg
(article number LF134)
(article number LF132)