8. Faults
8.1 Operating messages
Measuring range exceeded or excessive
Hi
turbidity.
Result below lowest limit of the measuring
Lo
range.
Replace batteries, no further tests possible.
Battery capacity is too low for the display
btLo
backlight, tests are still possible.
Store
Date
Cal
A user-calibrated method is indicated by an
RESULT
arrow while the test result is displayed.
Time
Cal
See section
6.8.5 Resetting to factory
calibration.
8.2 List of faults/error codes
Light absorption too great. Reasons: dirty optics,
E27/E28/E29
etc.
E 10 / E 11
Calibration factor out of range.
E 20 / E 21
Too much light reaching the detector.
E23/E24/E25
Too much light reaching the detector.
Battery capacity was too low during the
E 22
measurement. Change battery.
E 70
CL 6:
Factory calibration incorrect/deleted
E 71
CL 6:
User calibration incorrect/deleted
E 76
pH:
Factory calibration incorrect/deleted
E 77
pH:
User calibration incorrect/deleted
E 132
Prt:
Communication with DIT-IR failed
9. Spare parts
Description
Round vial, ∅24, with cap and
gasket
Plastic stirring rod,
length 13 cm
Brush, length 11 cm
10. Disposal
This product or parts of it must be disposed of in an
environmentally sound way. Use appropriate waste collection
services. If this is not possible, contact the nearest Grundfos
company or service workshop.
The 2006/66/EC guideline requires users to return all
Note
used and worn-out batteries and accumulators. They
must not be disposed of in normal domestic waste.
14
Quantity
Product No
Pack of 5
95727768
Pack of 12
95727769
1 piece
95727771
1 piece
95727772