Precaution
®
- Before using the Rightest
System to test your blood glucose, please read all the instructions and practice the tests
including quality control test. ( Refer to page 48.)
- Please do the quality control test regularly to make sure the test results are accurate .
®
- The Rightest
Blood Glucose Meter can only be used with the Rightest
brand Strips should not be used under any circumstances. The use of other brands of strips may give inaccurate results.
®
- The Rightest
Blood Glucose Monitoring System is intended for in vitro diagnostic use only. The testing result is
calibrated to be plasma equivalent for test with fresh capillary whole blood samples from the fingertip, palm and forearm.
®
- The Rightest
Blood Glucose Monitoring System is intended for self-testing. It should not be
used to diagnose diabetes mellitus.
®
- Rightest
Blood Glucose Monitoring System has not been validated for use on neonates.
Therefore, it's not intended for use of neonates.
®
- Rightest
Blood Glucose Meter hasn't been used for arterial blood testing yet, so it is still not suitable to be used for
arterial blood testing so far.
- Do the test at least 30 minutes after while entering another site with different ambient temperature.
- Pay attention to the environment protection when disposal of batteries.
- Using appliance in tropical and / or temperate climate.
4
®
Blood Glucose Test Strips. The use of any other
- Not to be exposed dripping or splashing.
®
- The minimum blood sample size of Rightest
Glucose Monitoring System is 1.4µL :(
1.0
µL
1.4
µL
Sample Size Example
We suggest you to take 1.4~2.5µL to do the test on glucose monitoring system. Blood sample size above 4.0µL is too
much which will contaminate the Smart Code Key.
Blood sample size below 1.4µL might cause inaccurate test result or might not start on meter measurement. In this case,
repeat the test with a new test strip.
PRECAUTOIN
®
The Rightest
Blood Glucose Meter GM300 will be set to mg/dL by default when sold in the United States.
When your display shows " mmol/L " during setting or test, please contact customer service. Use of the wrong
unit of measure may cause you to incorrect treatment.
Precaution
)
2.0
µL
3.0
µL
4
.0
µL
5