Error Messages and Trouble Shooting

About Temperature Error
1. In order to get accurate testing, please do testing between 10~40℃ (50~104℉ ).
2. When the ambient temperature is 0~9℃ (32~48℉) or 41~50℃ (106~122℉) , the "
symbol will blinking, you still can do the test but the result obtained is only for reference because the
test result under these range of temperature might not be correct. Repeat the test at an area with
temperature between operating range. ( 10~40℃ or 50~104℉ )
3. When you move from the area with temperature outside the operating range of test strip to another
area with temperature under operating range of test strip, please wait for 30 minutes before you do
the test.
4. When the temperature is below 0℃ ( 32℉ ) or over 50℃ ( 122℉ ), Meter can not do the test and the
"
" symbol will blink in this condition.
Please move the meter to environment with temperature between 10~40℃
( 50~104℉ ) and redo the test after 30 minutes.
58
About Battery Error
1. The "
can. You can still do the test.
2. The "
" warning
Please change the batteries immediately.
3. After changing the batteries you should do the quality control test.
About Check Key Error
When you see "
error message appears again, it would be Meter or Check Key defect.
Please return your Meter and Check Key to the store where you bought
About Smart Code Key Error
The Smart Code Key is faulty or not assembled well to the meter. Reinsert the Smart Code Key. If the error
screen appears again, contact Bionime Customer Service.
Error Messages and Trouble Shooting
" symbol is blinking when the battery power is low. Please change batteries as soon as you
" and "
" symbol are blinking when the battery is too low. Meter can not do the strip test.
" and "
" symbols showing on screen, please do the check key test again. If this
59