Performing a Test in Order
3. You can start measurement by the blood sample.
®
4. The minimum blood sample size of Rightest
Glucose Monitoring System is 1.4µL :( )
1.0
1.4
µL
µL
Sample Size Example
We suggest you to take 1.4~2.5µL to do the test on glucose monitoring system. Blood sample size above 4.0µL is too
much which will contaminate the Smart Code Key.
PRECAUTION
- Don't take too much blood sample, such as 4.0µl ( please see how much it is in the example picture ) or more,
to do the test. Too much blood might contaminate the Smart Code Key.
- Please clean it up right away if your Smart Code Key gets blood stain or moisture.
( Refer to page 55 for cleaning Smart Code Key )
- Blood sample size below 1.4µl might cause inaccurate test result or might not start on meter measurement. In
this case, repeat the test with a new test strip.
38
2.0
3.0
4
.0
µL
µL
µL
Applying blood sample to test strip
1. Touch the blood drop to the sample entry of the strip. The blood will be drawn into the strip
automatically.
2. Hold your finger on the sample entry until you hear a " beep " and the view window is totally filled with
blood. If the view window is not totally filled with blood and/or the test does not start, discard the test
strip and repeat the test with a new test strip. When you repeat the test, you may want to obtain the
blood sample from a different finger if the first attempt did not produce enough blood sample.
PRECAUTION
- When you do a test, please pay attention not to squeeze too much blood to pollute Smart Code Key.
Performing a Test in Order
39