When using, shall not touch battery and the patient
simultaneously.
Do not throw batteries into re.
The typical service life of the new and unused batteries is
300 measurements for the operation time is 60s.

Product errors and troubleshooting

Problem
Item to detect
· Battery depleted
· Batteries have been
installed with the
wrong polarity
· Batteries are not
installed properly
Unable to carry out
·
measurement as
current battery
capacity is too low.
· Measurement
distance too long
Target temperature is
·
beyond range of
measurements 89.6℉-
109.2℉(32℃-42.9℃)
·
Environment temperature
either exceeds the design
range 59℉-104℉(15℃-
40℃) or is unstable
·
Current state: All
symbols are ashing
on the screen. The
product is not usable.
·
Current battery
capacity is too low.

Body temperature

Body temperature varies from person to person and
uctuates during the course of the day. For this reason, it is
suggested to know one's normal, healthy forehead
temperature to correctly determine the temperature.
Body temperature runs approximately from 95.9°F to100°F
(35.5℃-37.8℃). To determine if one has a fever, compare the
temperature detected with the person's normal temperature.
A rise over the reference body temperature of 1°F(1℃) or
more is generally an indication of fever.
Di erent measurement sites(rectal, axillary, oral, frontal,
auricular) will give di erent readings. Therefore it is wrong to
compare the measurement taken from di erent sites.
Solution
Replace old
batteries with
new ones
Take out the
batteries and
re-install them
correctly.
Replace the
batteries
Follow the
instruction
manual and
repeat the
measurements
Please contact
customer
services.
Replace the
battery as soon
as possible.
6
The following are typical temperatures for adults, based
on di erent measurement sites:
Rectal
97.9°F-99.1°F
Axillary
94.5°F-99.1°F
Oral
95.9°F-99.5°F
Auricular
96.4°F-100.4°F

Care and Cleaning

1. The probe (comprising a temperature sensor and a distance
sensor) is the most intricate part of the thermometer, and
should be kept clean and intact to acquire accurate
readings.
Use the following method to clean the probe:
Gently swab the surface of the probe using a cotton bud
soaked in > 75% medical alcohol.
2. If the probe (sensor) is broken, please contact customer
services.
3. Use a piece of soft, dry cloth to clean the display screen and
external surface of the thermometer. If the thermometer is
very dirty, the cloth can be moistened with some medical
alcohol to clean the device.
4. The product is not waterproof. Do not clean the device with
detergents. Do not soak the thermometer in water or other
liquids.

Maintenance

1. This company has not authorized any agency or individual
to carry out product repairs or maintenance. Do not
attempt to disassemble or modify the thermometer if you
suspect functional issues with the device.
2. The IR thermometer is an extremely precise instrument. Any
improper maintenance, disassembly, or modi cation may
lead to inaccuracies of the product measurements.
3. If you suspect any product issues during the warranty
period, please contact customer services for subsequent
handling.

Disposal

Dispose of the battery in accordance with the
regulation applicable at the place of operation.
Dispose of at public collection point in the EU
countries – 2002/96/EC WEEE Directive.
If you have any queries, please refer to the local
authorities responsible for waste disposal.
Notes
• Please act according to local laws for disposal of used
batteries.
• Take out the battery if you are not going to use the unit for
a long time.
To protect the environment, dispose of empty
batteries at appropriate collection sites according to
national or local regulations.
Dispose of at public collection point in the EU
countries – 2006/66/EC Directive.
36.6℃ to 38℃
34.7℃ to 37.3℃
35.5℃ to 37.5℃
35.8℃ to 38℃