Overall Description; Product Performance - iHealth PT3 Manual De Usuario

Idiomas disponibles

Idiomas disponibles

measurement accuracy. We recommend taking only up to
3 continuous readings within a short period.
10. Do not directly face the sun or an air outlet of an air
conditioning or radiator device during the measurement
as this will cause changes to the forehead temperature.
Measurements should be taken in a stable environment
where possible.
11. Do not measure body temperature in an environment
with strong EM interference (examples include places
close to a working microwave, induction cooker, or
cellphone in-use) as EM interference may cause errors in
the reading or even device failure.
12. This product should be considered a personal device.
Clean and sanitize the product properly to prevent cross
contamination.
13. To clean a dirty thermometer probe, gently swipe the
probe using a cotton swab dipped in 75% alcohol. Let the
cleaned thermometer sit for at least 15 minutes before
taking more measurements.
14. This infrared thermometer meets requirements
established in ASTM Standard (E1965-98) Except of clause
5.2.2. It displays subject's temperature over a range of 89.6
℉~109.2℉. Full responsibility for the conformance of this
product to the standard is assumed by (manufacturer:
Famidoc Technology Co., Ltd. Add:
No. 212 Yilong Road, Hexi Industrial Zone, Jinxia, Changan
Town, Dongguan 523853, Guangdong Province, China.)
(2) About the product
1. This product is a precision device. Return the product to its
original packaging for proper storage after use. To ensure
accurate measurements, avoid the device or probe
contacting any liquid or droplets. Avoid tiny particles (such
as dust or powder) falling into the probe.
2. Avoid dropping or subjecting the product to external
forces. Do not disassemble or re-assemble the product on
your own.
3. Do not directly touch the probe with your ngers or blow
on it. Measurements taken using a damaged or dirty IR
probe may be inaccurate.
4. Keep the product at a place inaccessible to children to
prevent children from swallowing the batteries or small
parts.
5. Do not throw the thermometer or batteries into re to
prevent explosions.
6. Remove the batteries from the thermometer if the device
will not be used for more than one month.
7.If you are allergic to plastic/rubber, please don't use this
device.
8.The materials (ABS, TPU, metal) of expect contact with
patient had passed the ISO 10993-5 and ISO 10993-10
standards test, no toxicity, allergy and irritation reaction.
They are in compliance with the MDD requirements.
Based on the current science and technology, other
potential allergic reactions are unknown.

Overall Description

The thermometer is mainly comprised of a plastic casing, IR
2
temperature sensor, MCU, vibration motor, LED display screen
and batteries.
LED display
area
Measurement
button
LED screen instructions
Temperature
display area
Low power
reminder
Device dimensions: 5.39 in x 1.34 in x 1.54 in
Screen dimensions: 1.01 in x 0.75 in
Product weight: 90 g

Product performance

1. Measurement position: Centre of the forehead surface
2. Measurement distance: ≤1.18in(3 cm)
3. Power source: DC 3V; 2 x 1.5V AAA batteries
4. Measurement range: 89.6℉-109.2℉(32℃-42.9℃)
5. Measurement precision: ±0.4℉ within 95℉-107.6℉,
and±0.5℉ for other temperature ranges.
6. Resolution: 0.1℉
7. Clinical reproducibility: Within ±0.5℉
8. Operating conditions:
Temperature: 59℉-104℉(15℃-40℃)
Humidity: 15-85%RH, non-condensing
Atmospheric Pressure: 70~106kPa
9. Operation mode:
Adjusted mode: forehead
Direct mode: surface mode
10. Expected service life: 2 years
11. Reference Body Site: Oral
12. Software version: V1.0
ASTM laboratory accuracy requirements in the display range
of 98°F to 102°F(37℃) for IR thermometers is ±0.4°F(±0.2℃),
whereas for mercury in-glass and electronic thermometers,
the requirement per ASTM Standards E667-86 and E1112-86
is ±0.2°F(±0.1℃).
Thermometer
probe
Distance
probe
Battery cover
Celsius
Fahrenheit
3