"HOLD" key once and then your reading will jump over 100 pieces of record forward, when there
isn't over 100 pieces of record after current record, press this key to return to the 1st record.
- The luxometer can store 2,044 pieces of records at most.
4. Restore to factory defaults:
- Long press "MENU" key to enter function setting, and then short press this key to select restore
to factory defaults, and "DEF?" sign will be displayed on the LCD and flashes, at this moment you
can press "HOLD" to restore to factory defaults; press "MENU" or "REC/ERASE" key, you can exit
restore to factory defaults.
- After restoring to factory defaults, the defaulting state of the system is: USB.0 (no USB
transmission); APO1 (auto power off); 60S (auto record interval is set up as 60s); clearing all data
records.
5. Illuminance measuring
- In process of measuring illuminance, the sensitive mirror must be perpendicular to the
measured illuminating source to ensure accuracy of measured data..
- If distance between measured illuminating source and illuminometer, positions or different of
them, or the environment is different, then the measured data will be different; cause, the intensity
of light varies with change of distance, position, angle and environment. Since the intensity
of light will be varied with its distance, position, angle and surroundings to illuminometer,
measured date will be different accordingly.
- Make sure there is no scratch or dirt on sensitive mirror of luxometer, or the measured data
won't be accurate.
- When the illuminance exceeds the measuring range of the luxometer, "OL" sign will be
displayed.
- After completing measuring, cover the sensitive mirror with cap to avoid scratch on sensitive
mirror and keep it clean.
1. General specification
- LCD: 3 1/2-bit display, 1999-bit at most
- Over range display: "OL"
- Low voltage display:
- Type of sensor: digital visible light sensor
- Sampling rate: 100 times/s
- Power supply: one 1604A 6F22 9V battery
- Battery life: 200 hours for typical battery (alkaline battery)
- Dimensions: 195mmx 45mmx 26m
- Weight: about 185g (including battery)
2. Environment limit
- Indoor use
- Safety standards: EN61326: 2006;
EN55022: 1998+A1+A2;
EN55024: 1998+A1+A2
- Pollution grade: grade 2
VI SPECIFICATIONS
13