tempmate M1 Manual Del Usuario página 10

Ocultar thumbs Ver también para M1:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Manufacturer's Cer ficate of Valida on
 
Cer ficate No
VC150400013
imec Messtechnik GmbH cer fies that the products men oned on the following page(s) have been thoroughly tested,
validated and met performance accuracy specifica ons over the stated ranges.
Reference Instrumenta on
Device
Constant Temperature
Trough of Low
Temperature
Low‐Temperature
checking Equipment
All reference instruments are calibrated by SuZhou LonBo Calibra on and Tes ng Co.,LTD, a CNAS (China Na onal Accredita on Service)
accredited laboratory, and traceable to NIST according to Mul ‐Recogni on Agreement.
Environment Condi ons
 Air Temperature   
20.5°C
 Rela ve Humidity 
53.0%
Product Informa on
Product Name
tempmate‐M1
Inspector:
This cer ficate is valid for a period of 1 (one) year from date of sale.
Date
April 17 , 2015
Model
Calibra on Date
HC‐2010
March 28,2015
ATL60
March 28,2015
Valida on Informa on
Valida on Points
+60°C
+40°C
+25°C
0°C
‐5°C
‐25°C
Serial Number
TMM150400012
Accuracy/Vola lity
±0.05°C
±0.015°C
Tolerance
±0.5°C
±0.5°C
±0.5°C
±0.5°C
±0.5°C
±0.5°C
Produc on
Date
04/2015
Page                
1 of 1
Serial Number
JYDQ‐135
JYDQ‐136
Result
PASS
PASS
PASS
PASS
PASS
PASS
Expiry
Date
04/2016
V01/2015