Clinical accuracy validation method

TS9 is an adjusted mode clinical thermometer
the validated information for clinical accuracy in each adjusted mode
are:
GroupA1:
GroupA2:
GroupB :
GroupC :
cb: CLINICAL BIAS
L
: LlMITS OF AGREEMENT
A
σr: CLINICAL REPEATABILITY

Care and Maintenance

Lens/ Measurement Sensor:
Gently clean with an alcohol swab. Do not use water to wash the
thermometer lens directly.
‧ Thermometer:
Clean with a soft, dry cloth. Do not use water to rinse the device.
∆cb=-0.01°C, L
∆cb= 0.06°C, L
A
∆cb=-0.01°C, L
∆cb=-0.01°C, L
=0.18, σ
=±0.08°C
A
r
=0.22, σ
=±0.08°C
r
=0.20, σ
=±0.07°C
A
r
=0.18, σ
=±0.07°C
A
r