ELECTROMAGNETIC COMPATIBILITY
Guidance and manufacturer's declaration - electromagnetic immunity
The TS42S IR Thermometer is intended for use in the electromagnetic environment specified
below. The customer or the user of the TS42S IR Thermometer should assure that it is used in
such an environment.
Emissions
IEC 60601 Compliance Electromagnetic
Test
test level level
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2,5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predict-
ed theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in
which the TS42S IR Thermometer is used exceeds the applicable RF compliance level above, the TS42S
IR Thermometer should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the TS42S IR Thermometer.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Table 3 EMISSIONS
3 V/m
Environment Guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the TS42S IR
Thermometer, including cables, than the recommend-
ed separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d =
1
2 .
80 MHz to 800 MHz
d =
1
2 .
800 MHz to 2,5 GHz
d = 2.3
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m). Field strengths from fixed RF
transmitters, as determined by an electromagnetic
site survey,
should be less than the compliance level
a
in each frequency range.
the vicinity of equipment marked with the following
symbol:
Interference may occur in
b
English • 25