Wellell Domus 2+ Manual De Instrucciones página 19

Sistema de redistribución de presión alterna
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 21
Domus 2+ / English
3 V/m
MHz
GHz
80 % AM at 1
kHz
Radiated
RF
EM
385-6000
Fields
MHz, 9-28V/m,
IEC61000-4-3
80% AM(1kHz)
pulse
and other mod-
ulation
NOTE 1: U
is the a.c. mains voltage prior to the application of the test level
T
NOTE 2: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 3: These guidelines may not apply in all situations. Electromagnetic propagation is af-
fected by absorption and reflection from structures, objects and people
a)Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be pre-
dicted theoretically with accuracy. To assess the electromagnetic environment due to fixed
RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the device is used exceeds the applicable RF compliance
level above, the device should be observed to verify normal operation. If abnormal perfor-
mance is observed, additional measures may be necessary, such as reorienting or relocat-
ing the device.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
80
to
2.7
10 V/m 80 MHz to
2,7 GHz
80 % AM at 1 kHz
385-6000 MHz, 9-
28V/m,
AM(1kHz)
mode
mode
and
modulation
Recommended separation dis-
tance
d =
=
d
800MHz
=
d
2.7GHz
Where P is the maximum out-
put power rating of the trans-
mitter in watts (W) according
to the transmitter manufac-
turer and d is the recom-
mended separation distance in
10V/m
meters (m).
80%
Field strengths from fixed RF
pulse
transmitters, as determined by
other
an electromagnetic site sur-
vey ,
compliance level in each fre-
quency ranged.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
150kHz to 80MHz
P
80MHz to
0
6 ,
P
800 MHz to
1
2 ,
P
b
a
should be less than the
16
loading