Refer to the next page should you experience any errors during the calibration procedure.
Calibration Errors
ERROR
NO.
Err 1
Mathematical mistake; division by zero.
Cannot calculate calibration factor
Err 2
Weight loaded too low. Cannot
calibrate
Err 3
Unstable weight. Cannot calibrate
Err 5
Calibration disabled
Err 6
High saturation state (high A/D)
Err 7
Low saturation state (low A/D)
P/N UM600KL_600KG_Rev20201201_Web
DEFINITION
CHECK/REPLACE
Will show when trying to calibrate with no
calibration weight
While in calibration procedure add more weight in
load mode
Check the load cells mechanical surroundings
and see nothing touches them and the cables are
well connected
On the back of the display, locate the silver
extended screw and plastic tab. Use a nut driver
or pliers to remove the tab and gain access to the
push button underneath. Press and release one
time to enable proper calibration. When finished,
toggle the calibration switch to enable
The load cell cable is not connected correctly.
Check the connection, cables, and mechanical
connection.
See Err 6