Always remove the rechargeable batteries if you do not
intend to use the device for a week or more.
Batteries cannot be charged in the blood pressure monitor.
Recharge batteries in an external charger and observe the
information regarding charging, care and durability.
7. Error messages
If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «ERR 3», is displayed.
Error
Description
«ERR 1»
Signal too weak
AQ
«ERR 2»
Error signal
AP-B
«ERR 3»
Abnormal cuff pres-
sure
AP-C
«ERR 5» Abnormal result
«HI»
Pulse or cuff pres-
sure too high
«LO»
Pulse too low
Thuasne BP 3UG1-3F
Potential cause and remedy
The pulse signals on the cuff are
too weak. Re-position the cuff
and repeat the measurement.*
During the measurement, error
signals were detected by the
cuff, caused for instance by
movement or muscle tension.
Repeat the measurement,
keeping your arm still.
An adequate pressure cannot
be generated in the cuff. A leak
may have occurred. Check that
the cuff is correctly connected
and is not too loose. Replace the
batteries if necessary. Repeat
the measurement.
The measuring signals are inac-
curate and no result can there-
fore be displayed. Read through
the checklist for performing reli-
able measurements and then
repeat the measurement.*
The pressure in the cuff is too
high (over 299mmHg) OR the
pulse is too high (over 199 beats
per minute). Relax for 5 minutes
and repeat the measurement.*
The pulse is too low (less than
40 beats per minute). Repeat
the measurement.*
* Please immediately consult your doctor, if this or any other
problem occurs repeatedly.
If you think the results are unusual, please read through the
information in «Section 1.» carefully.
8. Safety, care, accuracy test and disposal
m
Safety and protection
 Follow instructions for use. This document provides important
product operation and safety information regarding this device.
Please read this document thoroughly before using the device
and keep for future reference.
 This device may only be used for the purposes described in
these instructions. The manufacturer cannot be held liable for
damage caused by incorrect application.
 This device comprises sensitive components and must be
treated with caution. Observe the storage and operating condi-
tions described in the «Technical specifications» section.
 The cuffs are sensitive and must be handled with care.
 Do not exchange or use any other kind of cuff or cuff connector
for measuring with this device.
 Only pump up the cuff once fitted.
 Do not use this device if you think it is damaged or notice
anything unusual.
 Never open this device.
 Read the additional safety information provided within the indi-
vidual sections of this instruction manual.
 The measurement results given by this device is not a diag-
nosis. It is not replacing the need for the consultation of a physi-
cian, especially if not matching the patient's symptoms. Do not
rely on the measurement result only, always consider other
potentially occurring symptoms and the patient's feedback.
Calling a doctor or an ambulance is advised if needed.
Ensure that children do not use this device unsupervised;
some parts are small enough to be swallowed. Be aware of
the risk of strangulation in case this device is supplied with
cables or tubes.