ERROR CODES
The Intelect Transport Combo displays error messages to inform the user of problems or potential problems with the unit, modality, or
accessories. These are numbered so the user can possibly correct the problem without the aid of service personnel. Use the following
Troubleshooting Charts to define the error codes, and locate the probable cause and possible remedies before contacting the dealer or
factory for technical service.
Code
Type
Number
Message
104
Message
105
Message
107
Warning
108
Warning
109
Message
200-399
Message
In the event that an Error message or Warning appears beginning with a 2 or 3, immediately stop all use of the unit and contact the dealer
or Chattanooga Group for service. Errors and Warnings in these categories indicate an internal problem with the unit that must be tested
by Chattanooga Group or a Field Service Technician certified by Chattanooga Group before any further operation or use of the system.
Use of a unit that indicates an Error or Warning in these categories may pose a risk of injury to the patient, user, or extensive internal
damage to the system.
TROUBLESHOOTING ERROR MESSAGES
Troubleshooting Error Messages
The following messages are displayed on the LCD under the following conditions:
Message
Over Temp
Unplugged
Uncoupled
Probable Cause
User attempted to perform an
electrotherapy session, but both channels
are already in use.
User selected a two channel
electrotherapy treatment, but at least one
channel is already in use.
Bad Contact Quality
Shorted Lead Wires
While performing ultrasound therapy
2
using the 10 cm
sound at an intensity
2
greater than 15 W (1.7 W/cm
), user
attempted to begin an electrotherapy
session on a second channel.
Refer to Warning on next page.
WARNING
Displayed When
sound head reaches a temperature which could damage the crystal
sound head not plugged in or faulty sound head
sound head is not making good contact with the patient
TROUBLESHOOTING
Possible Remedies
A. Wait until the previous therapy session finishes.
B. Press the STOP button to end the therapy session on
either channel.
A. Wait until the previous therapy session finishes.
B. Press the STOP button to end the therapy session on
either channel.
A. Make certain Electrodes are making proper contact with
the treatment area.
B. Make certain Lead Wires are properly connected to
Electrodes.
C. Replace Electrodes and Lead Wires.
A. Check Electrodes and Lead Wires. Make certain Lead
Wires are not damaged and are properly connected
to the system. Make certain Lead Wires are properly
connected to the Electrodes and that electrodes are
not damaged and are making proper contact with
treatment area.
B. Replace Lead Wires and Electrodes.
A. Use a smaller sound head.
B. Set the ultrasound at an intensity less than
2
15 W (1.7 W/cm
).
C. Wait until ultrasound session is complete.
Refer to Warning on next page.
39