SYSTEM
WAVEFORMS (CONTINUED)
TENS- SYMMETRICAL BIPHASIC
The Symmetrical Biphasic waveform has a short pulse duration and is
capable of strong stimulation of nerve fibers in the skin and in muscle. This
waveform is often used in portable muscle stimulation units, and some
TENS devices.
Output Mode
Available on Channel
Treatment Time (Stim)
Treatment Time (Combo)
Mode Selection (Stim)
Mode Selection (Combo)
Output Intensity
Amplitude Modulation
Burst Frequency
Cycle Time
Frequency
Frequency Sweep
Phase Duration
Ramp
Sweep Time
Sweep Low Frequency
Sweep High Frequency
IRMS
DC component
GALVANIC: CONTINUOUS
Galvanic Current is a direct current flowing in one direction only.
The current can be continuous or interrupted.
Output Mode
Available on Channels
Treatment Time
Mode Selection
Output Intensity
Cycle Time
Polarity Reversal
With Polarity Reversal On, Polarity will change in
IRMS
DC component
47979 INTELECT TRANSPORT 2 COMBO (4778) AND ULTRASOUND (4776) USER MANUAL
Electrodes
1, 2
1-60 min
1-30 min
CC or CV
CV
0-140 mA (CC) 0-140 V (CV)
0% (off) to 100% on 10% steps
0-10 bps
Continuous or User Defined
1-200 pps
On/Off
30-400 sec
0-5 sec
14sec
1-199 pps
2-200 pps
0-50 mA
No
Electrodes
1, 2
1-60 min
CC
0-40 mA (CC)
Continuous, or User Defined
On or Off
the middle of the treatment time.
0-44 mA
Yes
TRÄBERT (ULTRAREIZ)
The Träbert Current is a monophasic waveform with a phase duration
of 2 ms and a pause of 5 ms resulting in a frequency of approximately
143 Hz.
Output Mode
Available on Channels
Treatment Time
Mode Selection
Output Intensity
Frequency
Polarity Reversal
With Polarity Reversal On, Polarity will change
Phase Duration
IRMS
DC component
GALVANIC: INTERRUPTED
Galvanic Current is a direct current flowing in one direction only.
The current can be continuous or interrupted.
Output Mode
Available on Channels
Treatment Time
Mode Selection
Output Intensity
Pulse Duration
Phase Interval
Polarity Reversal
change in the middle of the treatment time.
Polarity Reversal Ramp
IRMS
DC component
EN
Electrodes
1-60 min
0-80 mA (CC)
143 pps
On or Off
in the middle of the treatment time.
2 ms
0-47mA
Electrodes
1-60 min
0-40 mA (CC)
136 sec
25 usec
On or Off
With Polarity Reversal On, Polarity will
1 sec
0-41 mA
1, 2
CC
Yes
1, 2
CC
Yes