RS Pro RS-9331 Manual De Instrucciones página 4

Comprobador eléctrico
Idiomas disponibles
  • ES

Idiomas disponibles

Electrical Tester / English
2. Appropriate Usage
The instrument may only be used under those conditions and for those purposes
for which it was conceived. For this reason, in particular the safety references,
the technical data including environmental conditions and the usage in dry
environments must be followed.
When modifying or changing the instrument, the operational safety is no longer
ensured.
The instrument may only be opened by an authorized service technician.
The voltage detectors are designed to be used by skilled persons and in accordance
with safe methods of work.
Before using a voltage detector with audible indicator at locations with a high
back ground noise level, it has to be determined whether the audible signal is
perceptible.
3. Specifications
LEDS
LED voltage range
12V to 1000V AC/DC
LED resolution
±12,24,50,120,230, 400, 690,1000V AC/DC
Tolerances
Complies with EN 61243-3:2014
Frequency range
0/40Hz to 400Hz
Response time
≤1 second
Auto power on ≥12V AC/DC
Auto power on
LCD
1999 counts (3 1/2 digit ) LCD display with
LCD display
bargraph & backlight
Voltage range
6 V to 1000 V AC/DC
Resolution
1V AC/DC
±3.0% of reading ±5 digit
Tolerances
Frequency range
0/40Hz to 400Hz
≤1second
Response time
Auto power on ≥ 6V AC/DC
Auto power on
4
Voltage detection
automatic
Full range
Polarity detection
Range detection
Automatic
Maximum 3.5 mA at 1000 V
Internal basic load
350 kΩ / Is<3.5 mA (no RCD tripping)
impedance Peak current
Operation time
Duration Time = 30 seconds
Recovery time
Recovery Time = 240 seconds
Switchable Load
~7KΩ
Is (load)=150mA
Peak current
RCD tripping I
~30mA@230V
Continuity Test
0 to 400 kΩ
Accuracy
nominal resistance ±50 %
≤5 µA
Test current
Resistance measurement
0Ω to1999Ω
1 Ω
Resolution
±(5 % rdg +10 digits) @ 20 °C
Tolerance
±5 digits / 10 K
Temperature coefficient
≤30 μA
Test current
Single-pole Phase Test
100V to 1000V AC
Frequency range
50 Hz to 400 Hz
Rotary Field Indication
Voltage range (LEDs)
100...1000V
Frequency range
50/60Hz
Measurement principle
Double-pole and contact electrode
Safety standards
EN61243-3:2014
Agency approvals
TÜV Rheinland -GS
Overvoltage protection
1000V AC/DC
Measurement category
CATⅢ 1000V/CATⅣ 600V
Protection degree
IP64
Power supply
2x1.5V "AAA" Batteries
Power consumption
max.30mA / approx. 250mW
Temperature range
-10 C up to 55 C
o
Humidity
max.85% relative humidity
Electrical Tester / English
o
5
loading

Este manual también es adecuado para:

193-8692