Operation
68
Zero adjustment
Please insert zero cell (distilled water)
or press <Start/Enter>
Zero cell
(H O dist.)
2
inner
turn-up lid
Zero cell
(H O dist.)
2
The zero adjustment window pops
04/16/07 9:52
up.
3 Close the inner turn-up lid.
4 Depending on the cell type, insert
the zero cell as follows:
Round cell:
Insert the round cell in the round
cell shaft so it touches the bottom.
If the inner turn-up lid is opened
too wide, a message prompts you
to close the inner turn-up lid.
Rectangular cell:
Open the inner turn-up lid.
Insert the rectangular cell
vertically so it touches the bottom
and left edge of the cell shaft. The
opaque sides of the rectangular
cell must point to the front and
back.
The photometer has an external
light recognition. If there is too
much external light, a message
prompts you to close the cell shaft
cover.
®
photoLab
6600 UV-VIS
ba75850e04
02/2011
loading