BGS technic 2193 Manual De Instrucciones página 8

TWO-POINT CALIBRATION
1. For the two-point calibration, please take the calibration plate (round metal disc) and the calibration
sheet (rectangular sheet labeled 3.9 mils, 99μm) at hand.
2. Switch on by pressing the trigger.
Press "CAL" button for 4 seconds to start the calibration.
Then the display flashes "- - 1".
3. Press the device with the probe on the calibration plate (round metal plate), without calibration
sheet , then press and hold the trigger till the display shows a reading.
4. The trigger can now released and the device can be removed from the calibration plate. Adjust with
the ▼ ▲ buttons to select the displayed value to Zero.
3. Press "CAL" button to confirm. The display first shows "- - -" and then "- - 2".
4. Press the device with the probe on the calibration film (3.9 mils, 99μm), the calibration plate (round
metal plate) must be lie under the calibration film ,press and hold the trigger till the item shows a
reading.
4. The trigger can now released and the device can be removed from the calibration plate.
Use the ▼ ▲ button to set the displayed reading on the thickness of the calibration plate (3.9 mils
or 99μm).
5. Press "CAL" button to confirm. The display flickers "- - -" and then "- - 3".
6. Hold the "CAL" button for 4 seconds to leave the calibration mode and the device turns off
automatically.
Note:
If you hold the "CAL" button for 4 seconds during calibration, the unit will return to the application
mode and NOT stores the calibration value!
OPERATING INSTRUCTIONS
1. Keep the instrument away from surfaces and electromagnetic fields when you put it into operation.
2. Press the trigger to activate the device and wait until the symbols "run" and "HOLD" will be
displayed.
3. Press the sensor firmly to the surface to be measured.
4. To measure, pull the trigger and immediately releasing it.
Do not take away the probe from underground before the measured value is displayed.
4. If the measured coating outside of the range, still displays a measured value. This is less accurate
than indicated in the specifications.
MAINTENANCE
Replacing the battery:
1. The device is powered by a 9 volt battery
(NEDA 1604, IEC 6F22)
2. Remove the battery cover carefully, the coverage
under a slight pressure on the arrow icon slowly slide down.
3. Remove the old battery and the battery cable away from
the battery housing. Insert the new battery.
Put the cover on the battery housing
Cleaning
Wipe regularly the housing of the unit with a damp cloth and
mild detergent.
Do not use acids, alkalis or solvents to clean.
SW-Stahl und Werkzeugvertriebs GmbH
L
e
v
e
k r
u
s
r e
t S
a r
s
s
e
6
5
D
4 -
2
8
9
7
R
e
m
s
c
h
e
d i
Tel. +49 (0) 2191 / 46438-0
F
ax +49 (0) 2191 / 46438-40
- E
M
a
: l i
n i
o f
@
s
w
s
a t
l h
d .
e
Attention
loading