SPECIFICATION
Model No.:
Type:
Measurement
Range:
Accuracy:
Power Supply:
Battery Life:
TROUBLE SHOOTING (1)
Abnormality
LCD shows Low Batter
symbol
The
unit
does
measure. Readings are
too high or too low.
An irregular heartbeat
symbol occurs.
Incorrect operation
BP-4715
-------------------------------------
Oscillometric; Automatic air
inflation by air pump and
automatic deflation
-------------------------------------
Pressure: 40~280mmHg
Pulse: 30~160 Pulses
-------------------------------------
Pressure: within ±5mmHg
Pulse: within ±5%
6V DC (4 "AA" batteries)
-------------------------------------
Approx.
250
(180mmHg,
once /day, 22°C)
Probable Reason
Batteries are low.
Pneumatic
blocked or cuff is too
tightly wrapped.
not
Pressure
unstable
measurement.
The cuff position is not
correct.
Irregular heartbeat
Some
interference
inflation
operation
measuring
Operation
Environment:
Storage
Environment:
Classification:
Cuff Size:
Memory:
Dimensions:
times
Weight:
Corrective action
Install new batteries.
system
Make certain the cuff is wrapped
around your arm correctly and re-
measure.
system
was
Measure again.
before
move or speak during measurement.
Sit comfortably and still.
the cuff is at the same level at your
heart.
Relax
measure again. If the symbol appears
again, consult your physician or other
health care professional..
in
or
wrong
Refer to the inflation step in "Taking
during
blood pressure" and process again.
Temperature: 5~40°C
Humidity: < 85%RH
----------------------------------------
Temperature: -20~60°C
Humidity: < 95%RH
Class II, type B
----------------------------------------
220mm ~ 480mm (+/- 5mm)
----------------------------------------
90 measurements including
date and time
----------------------------------
141mm (5.5inch)------ (L)
121mm (4.7inch) ------(W)
82mm (3.2inch) ------- (H)
----------------------------------------
340g (0.75 LB)
Stay calm.
Make sure
for
about
5
minutes
Do not
and
P-7