12. TECHNICAL DATA
12.1 Specifications
Product Description
Model
Suction
Vacuum Levels
Cycles Per Minute
Breast shield Size
Power Source
Battery Life
Additional Function
Unit Weight
Unit Dimensions
Operating Environment
Storage Environment
Transportation
Expected Service Life
Ingress Protection Rating
Classification
Protection Class
This appliance conforms to the following standards:
IEC 60601-1-11:2015 Medical electrical equipment —Part 1-11: General requirements for basic safety and essential
performance –Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the
home healthcare environment and complies with the requirements of IEC 60601-1-2 (EMC), and AAMI/IEC 60601-1 (Safety)
standards. The manufacturer is ISO 13485 certified.
Note: Suction pressure may be slightly lower than shown above if the pump is operated at higher altitudes or in an aircraft.
24
|
ENGLISH
Hygeia FIT Pro
The pump is capable of single and double pumping.
Stimulation mode: Dual-Kit: 40 to 100 mmHg
Stimulation mode: Single-Kit: 50 to 130 mmHg
Expression mode: Dual-Kit: 50 to 270 mmHg
Expression mode: Single-Kit: 50 to 210 mmHg
Stimulation mode: 6 Levels
Expression mode: 12 Levels
Stimulation mode: 80 to 120 C.P.M.
Expression mode: 20 to 65 C.P.M.
28mm
AC Adapter – Output: 5VDC, 1000mA.
Input: 100-240VAC, 50/60Hz, 0.25A.
Model # MPSUL0501000
1500 mAh Lithium-ion rechargeable battery
At least 2 hours after full charge cycle
Automatic Power-off after 45 min
Approx. 328.4 g (11.6oz)
110 x 93 x 55mm (L x W x H)
Temperature: 5
o
C to 40
o
Humidity: 15 to 93%RH
Pressure: 86kPa to 106kPa
Temperature: -20
C to 55
o
Humidity: 15 to 93%RH
Your device has been designed and developed with the greatest possible care and has an
expected service life of 300 hours.
IP22
Internal Powered Equipment, Type BF Applied Part
Class II/Double Insulated
Electric Breast Pump
C (41
o
F to 104
o
F)
C (-4
F to 131
F)
o
o
o
Specifications are subject to change without notice.
ART-70295-01-AA