Maintenance; Checkup; Cleaning And Replacement Of Parts - gc Labolight DUO Instrucciones De Uso

Unidad de fotopolimerización led dual para resinas de composite indirecta
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 118

6. Maintenance

6.1 Checkup

Perform checkup and maintenance by taking the following recommended intervals as an indication:
■ Be sure to perform regular checkup of the equipment and components.
Verifying LEDs being lit
Verifying state of the turntable
Daily checkup
Checking the lens cover for
contamination
Whether the connections are sound
Checking the step mode switch for
Weekly checkup
contamination
Checking inside the curing chamber
for contamination
Replacing the lens cover
(Replacement interval as an
Parts replacement
indication: 1 year *depends on
frequency of usage)
■ Be sure to clean the curing chamber regularly, as otherwise, the contaminated chamber could not exercise
suffi cient curing performance. If fl ammable disinfectant is used (e.g. Isopropyl alcohol based), there is risk
of ignition/explosion if lamp is started with vapors in the chamber.
● In case a LED unit fails, other LED units can still properly irradiate light. However, as the total amount of
light is reduced, the curing time should be extended and curing status should be suffi ciently monitored.
The failed LED unit should be promptly repaired.

6.2 Cleaning and Replacement of parts

Be sure to plug off the power cord so that the unit is not activated during cleaning or replacement work.
■ Maintenance, if intended after a prolonged use, should be conducted after the unit temperature suffi ciently
lowers.
CAUTION
Contents
After pressing the timer key with the door opened, verify
lighting of respective LEDs and proper rotation of the
turntable. With any failure existing, a curing defect occurs.
(For details, see Section 4.5, "Startup and checkup.")
Clean the lens cover if contaminated.
(See Section 6.2, "Cleaning and replacement of respective
portions.")
Verify the sound connection of the AC adapter and power
cord, and check them for accumulation of dust.
Clean the step mode switch if attached with dust and dirt.
(See Section 6.2, "Cleaning and replacement of respective
portions.")
Clean the curing chamber if contaminated.
(See Section 6.2, "Cleaning and replacement of respective
portions.")
Heavy discoloration, contamination or scratches of the lens
might result in a curing defect. When checkup of the lens
cover is performed and find discoloration, contamination or
scratches which cannot be cleaned, replace the lens cover.
* Lens cover potentially be discolored by aging.
(See Section 6.2, "Cleaning and replacement of respective
portions.")
CAUTION
Note
CAUTION
Details
EN 21
Tabla de contenido

Solución de problemas

loading

Tabla de contenido