Open the blister pack. Take out the filter/opener. Hold
the tip of the Freezpen pointing downward. Place
the filter/opener and make sure that it sits flat at the
bottom of the loading cavity. The trocar of the opener
must be pointed upward. Take the capsule and place
it in the cavity,on top of the filter/opener device.
Installation of the assembled filter/opener device.
Step 3
I
Puncture the capsule
Gently slide the capsule into the Freezpen's loading
cavity and start screwing the capsule turning it
clockwise in a quick and uninterrupted movement
until hand-tight.
10
Step 4
I
Use your Freezpen
Remove the applicator cap. Push the trigger to start
the flow of cryogen. Release the trigger to stop the
flow of cryogen. You may notice that after releasing
the trigger, the remaining gas in the applicator is
expelled. To switch the applicator, unscrew the current
one and screw the other one on the Freezpen.
The table below gives the estimated application
with a continuous flow, time depending
on the applicator and the cartridge size.
The measurements were done at 21°C (70°F).
Green
Orange (60mm)
White (60mm)
Freezpen 35g
Lesion
210-240s
4-8 mm
565-590s
1-4 mm
295-325s
2-6 mm