11. Notes about malfunctions
GB
Fault
No light curing
Handpiece becomes
warm
Polymerisation is
inadequate
Red LED (8) on
charging unit (7) is lit
The charging unit (7)
cannot be switched on
(green LED is not on)
In the event of faults or damage please call our free Service Hotline,
tel. 0 800-43 72 52 2 or contact your local fi eld sales representative.
In other countries please contact your appropriate Heraeus Kulzer GmbH subsidiary
or your local dealer.
12. Document history
01.01.2008
Possible cause
Battery not charged
Continuous duty for
30 min
Battery is down
Light exit window (1)
soiled
Overheating protection
system activated
Short circuit across the
charging contacts
The power supply cable
is not connected up
properly
Power supply cable is
damaged
The charger (7) is not
operating
First edition
Remedy
Charge battery as des-
cribed in chapter 7.1.
Switch off handpiece and
allow to cool down
Charge battery as des-
cribed in chapter 7.1.
Clean as described in
chapter 8. Cleaning.
Allow the base unit to
cool down
Battery faulty. Call the
Service Hotline or contact
your local field sales
representative.
Charging unit (7) faulty.
Call the Service Hotline
or contact your local field
sales representative.
Plug the cable into
the charger (7) and the
power socket
Change the power
supply cable
Call the Service Hotline
or contact your local field
sales representative.