FIREBOY eco – Operating instructions V00
3.2.4 Removing and changing the gas cartridge
W
ARNING
The gas cartridge may only be replaced if the gas cartridge adapter is
connected to the device and closed off, if possilbe.
The different gas cartridges are removed and changed as follows:
If the gas cartridge adapter is equipped with a milled knob, turn the knob
fully clockwise in the "OFF" direction.
Remove the empty cartridge.
Fit a new cartridge.
If the gas cartridge adapter is equipped with a milled knob, turn the knob
fully counterclockwise in the "ON" direction.
The gas cartridge has been changed.
C
AUTION
Gas will begin to flow.
3.2.5 Connecting to the central gas supply or to a gas cylinder
Connect the tubing from the central supply or from the gas cylinder to the
previously installed adapter and fix with a clip if necessary.
FIREBOY eco is now connected to the gas supply.
3.2.6 Setting the right flammable mixture
The GAS KNOB (gas) can be used to set the gas supply and therefore the
flame size while observing the flame itself. Rotating the knob clockwise
reduces the gas supply (-), rotating it counterclockwise increases it (+):
The air knob (air) can be used to regulate the air supply and therefore the
flame temperature. Rotating the knob clockwise reduces the air supply (-),
rotating it counterclockwise increases it (+).
10
Start-up
INTEGRA Biosciences