Error indication
Device errors which endanger the measuring process (red LED)
Error temperature
Error pump
Error memory
Error alcohol sensor
Error calibration
Dok.-Nr.:
059-07-11000300-k.doc
Possible cause
Temperature
The temperature
established in the device is
beyond the specified range
of between -5 and +50 °C.
Pump
An error was detected in
the sampling system.
Memory
An error has occurred
during reading or writing
access to the internal
memory.
Alcohol sensor
An error was detected on
the alcohol sensor.
Calibration
An incorrect calibration
date was found.
/ 07.08
AlcoQuant 6020
Remedy
Measurement may be faulty.
Beyond the specified temperature range
the measured values established may
be subject to greater error tolerance
than specified. The device must be
switched off and brought to within the
specified temperature range (by cooling
or heating).
Then the device must be switched on
again.
If the error occurs although the device is
evidently
within
temperature
range,
Service.
Switching the device on and off.
If the fault occurs again during the self-
test or when measuring, please inform
Service.
A faulty measurement cannot occur
because the measured value is rejected
in the case of "Faulty sample collection"
Measuring is not possible.
The measured value may not be
displayed correctly.
Data cannot be read correctly off the
memory and the established measured
values cannot be logged. Measure
again.
If the error recurs, please contact
Service.
Measuring is not possible.
Switch the device off and then on again.
If the error recurs, please contact
Service.
Measuring is not possible.
Device must be calibrated again, please
contact Service.
If
the
device
has
calibrated, it is possible that the current
date settings are not correct.
Here, too, please contact Service.
the
specified
please
contact
already
been
33