Invacare Medley Ergo Manual Del Usuario página 9

Ocultar thumbs Ver también para Medley Ergo:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 183
Electromagnetic interference between the bed and other electrical products can occur. To reduce or eliminate such
electromagnetic interference, increase the distance between the bed and the products or switch them off.
This medical bed can be used together with medical electrical equipment connected to the heart (intracardially) or
the blood vessels (intravascularly), provided that following points are respected:
• The medical bed should be equipped with means for potential equalization connection marked out by a symbol
shown in the back of this manual.
• Medical electrical equipment as describe above, should not be fixed on the bed's metallic accessories such as side
rails, lifting pole, drip rod, bed ends ect. In addition, the medical electrical equipment power supply cord should be kept
clear of the accessories or any other moving part of the bed.
Make sure that there are nothing under, over, or near the bed that can limit the movement of the bed or the mattress
support, such as furniture, window frames and storage boxes.
Always lower the bed to the lowest position before leaving the patient in the bed unattended.
There is a risk of entrapment of fingers in the bed moving parts.
Never lower deck height of bed more than 40 cm when using bed with a
lifter. Use ground clearance info in section 8 of this manual to determine
lifter compatibility with bed before use.
It can be dangerous to roll over the mains cable.
Do not bring mains cable into moving parts.
Disconnect the plug from the mains before moving the bed.
The cables must be mounted in such a way that they are
kept clear of the floor and do not block the castors.
We recommend to mount the mains cable on the hook for this purpose, see picture.
7
Tabla de contenido
loading

Tabla de contenido