USAGE
Cleaning and care
Clean with damp cloth and a mild soap. Dry with a cloth after cleaning
Environmental Conditions
The device is Waterproof.
A Waterproof device can be used in a wet or humid environment and
submerged in up to 1-meter-deep fresh water for a maximum of 30
minutes. No contact with salt water or chlorinated water is allowed.
Dry with a cloth after contact with fresh water or humidity. Clean with
fresh water in case of accidental exposure to other liquids, chemicals,
sand, dust, or dirt and dry with a cloth.
MAINTENANCE
The device and the overall prosthesis should be examined by a healthcare
professional. Interval should be determined based on patient activity.
Note: Do not disassemble the 90° Unity
components, as this may result in damage of internal sealings and thus
loss of elevated vacuum.
Caution: The locking pin should be exchanged at least once a year.
Check the release button for ease of use. If the release button is hard to
use, the mechanism is damaged. Exchange the device.
Check if the device is airtight. If the device is not airtight, clean it as
follows:
1. Examine the pin hole. If necessary, remove any dust, dirt, or similar
contamination.
2. Trap the pin hole with e.g., tape or plasticine.
3. Use low air pressure to clean the socket.
tube connector in its
®
13