Sensor's usage and offset
Prior to use:
1. Remove the blue rubber cap from the NeuLog
oxygen probe.
2. Unscrew the tip-end membrane cap of the probe,
inside you will see a metal electrode.
3. Add 1 mL (half of the cap) of the "DO Filling Solution"
directly into the tip-end which was just unscrewed.
4. Carefully reattach the tip-end of the probe by
screwing it back into the probe (do not over tighten).
5. Do not replace the blue rubber cap until after testing.
Although the NeuLog oxygen sensor does not need to be
calibrated prior to each use, it is recommended to offset
its probe.
Open air mode:
Offset of the probe assumes a standard atmospheric
oxygen concentration of 20.9% (assuming stable levels
in the Earth's atmosphere at sea level).
1. Connect the oxygen sensor to a voltage source
(USB-200, BAT-200, WIFI-201).
2. Wait about 5 minutes after connecting the sensor to
the voltage source before you offset the sensor. For a
more accurate measurement, look at the module
window and wait for the reading to stabilize (it could
take a little longer).
3. Press and hold the oxygen sensor's push button for 3
seconds.