Properly Aligning the Charger
1. Press the power button. The indicator light will come on again, and the Charger will
begin beeping steadily to signal that it is searching for the Stimulator.
2. Place the Charger on the chest in the area of the Stimulator.
3. Position the Charger over the Stimulator.
The beeping will stop when the Charger is partially aligned with the Stimulator,
but the Charger should be fully aligned with the Stimulator for optimal charging.
Charging time may increase if the Charger is not fully aligned with the Stimulator. To
fully align the Charger over the Stimulator, position the bottom curve of the Charger
approximately 1 cm lower than the bottom curve of the Stimulator, as shown below.
You may use the tips of your index and middle fingers to help locate the Stimulator
under the skin. Lay your fingers gently on the skin without applying pressure or
gripping. You may use this method to locate the center of the Stimulator.