WARNINGS CONTINUED:
Parts and Accessories section in this manual (page
20).
•
Never operate this product if any of the parts are not
working properly or have been damaged.
•
Do not use the Aeroneb
of a flammable anaesthetic mixture with air, oxygen or
nitrous oxide. As with any electrical equipment, take care
when using the device in a potentially oxygen enriched
environment.
•
Do not use the Aeroneb
breathing systems or lung ventilator breathing systems.
•
Do not place or drop this product into water or any liquid
when it is plugged into a power source.
•
Keep this product and cords away from heated surfaces.
•
Be sure the device has been properly cleaned before
use to avoid possible contamination.
•
If after proper cleaning it takes longer than 12 minutes to
deliver 3mL of normal saline (0.9%), the Nebulizer Unit
should be replaced.
•
The user should ensure that exhaled medication is
directed away from bystanders.
•
Do not microwave any parts.
•
By connecting the AC/DC adapter to the Aeroneb
Control Module, the Aeroneb
batteries are not being charged.
•
As with any cables keep all cables tidy to avoid tripping
or strangulation hazards and take particular care around
children.
•
Do not nebulize alcohol-based drugs in an oxygen-rich/
high pressure environment.
Go Nebulizer in the presence
®
Go Nebulizer with anaesthetic
®
Go Control Module and
®
Go
®
3