WARNINGS CONTINUED:
Parts and Accessories section in this manual (page
20).
Never operate this product if any of the parts are not
working properly or have been damaged.
Do not use the Aeroneb
of a flammable anaesthetic mixture with air, oxygen or
nitrous oxide. As with any electrical equipment, take care
when using the device in a potentially oxygen enriched
environment.
Do not use the Aeroneb
breathing systems or lung ventilator breathing systems.
Do not place or drop this product into water or any liquid
when it is plugged into a power source.
Keep this product and cords away from heated surfaces.
Be sure the device has been properly cleaned before
use to avoid possible contamination.
If after proper cleaning it takes longer than 12 minutes to
deliver 3mL of normal saline (0.9%), the Nebulizer Unit
should be replaced.
The user should ensure that exhaled medication is
directed away from bystanders.
Do not microwave any parts.
By connecting the AC/DC adapter to the Aeroneb
Control Module, the Aeroneb
batteries are not being charged.
As with any cables keep all cables tidy to avoid tripping
or strangulation hazards and take particular care around
children.
Do not nebulize alcohol-based drugs in an oxygen-rich/
high pressure environment.
Go Nebulizer in the presence
®
Go Nebulizer with anaesthetic
®
Go Control Module and
®
Go
®
3
loading