Página 3
Storage Temperature: -10°C to 60°C (14°F to 140°F); <80 % RH Altitude: Up to 2000 meters Pollution Degree: 2 Explanation of Symbols Attention refer to operation instructions. Approvals CONFORMS TO UL STD. 61010-1, 61010-2-032, 61010-2-033 CERTIFIED TO CSA STD. C22.2 #61010-1-12, 61010-2-032, 61010-2-033 3131884 KING INNOVATION...
Página 4
INSTRUCTION MANUAL 2. GENERAL SPECIFICATION The King Innovation KING CM Digital Milliamp Clamp Meter is ideal for use in troubleshooting two-wire irrigation, landscape lighting and electrical systems. The KING CM is for use in indoor and outdoor dry locations. Digital Display:...
Página 6
Range Resolution Accuracy 60.00V 0.01V 1.0% + 2dgts 600.0V 0.1V Input impedance: 2MΩ 3-5 Continuity Range Buzzer Function Ohm < 45Ω 3-6 Resistance (Ω) Range Resolution Accuracy 600.0Ω 0.1Ω 6.000KΩ 0.001KΩ 1.0% + 2dgts 60.00KΩ 0.01KΩ 600.0KΩ 0.1KΩ KING INNOVATION...
Página 7
AC measurement indication DC measurement indication Current measurement indication Voltage measurement indication ZERO ZERO indication HOLD Data HOLD indication PEAK PEAK hold indication Low pass filter indication Continuity test indication Measurement unit Ω Resistance measurement indication Measurement unit KING INNOVATION...
Página 8
3 Backlight Button 4 Clamp Opening Handle 5 Function Select Dial 6 ZERO Button 7 PEAK and Ω Buttons 8 Data HOLD & LPF Button 9 LCD Display 10 COM Input Terminal 11 Positive Input Terminal 12 Battery Cabinet KING INNOVATION...
Página 9
Press “ZERO” button to enter the “ZERO” mode, ZERO Indicator will appear and the display will ZERO. The reading is stored as reference value for the subsequent measurement. Press the “ZERO” button again, to exit the ZERO mode. KING INNOVATION...
Página 10
The measured imbalance current value is shown on the display. If the display is not giving a steady and consistent reading, press the HOLD button and read the display after the clamp is removed from the conductors. KING INNOVATION...
Página 11
MILLIAMP CLAMP METER Fig. 1 Normal Measurement: INCORRECT CORRECT Fig. 2 Measuring out of balance leakage current: LOAD LOAD 3-Phase 3-Wire System Single-Phase 2-Wire System KING INNOVATION...
Página 12
“+” terminal and the black lead to the “COM” terminal. Place the red and black test leads at both the positive and negative points to be measured. The display will show the value of the measurement being taken. KING INNOVATION...
Página 13
45 Ω. NOTE: When measuring continuity, all power to the circuit or cable being tested MUST be turned off to prevent damage to the user or the clamp meter. Short Circuit Open Circuit KING INNOVATION...
Página 14
The display will show the resistance value of the measurement being taken on the display. NOTE: When measuring resistance, the circuit should be powered off and all capacitors should be completely discharged prior to testing. KING INNOVATION...
Página 15
Never use the meter unless the battery cover is in place and the screw is tightened. 8-1 Cleaning: Use only a soft cloth dampened with water or neutral detergent for cleaning the meter. Do not use abrasives, solvents or harsh chemicals. Allow to dry thoroughly before use. KING INNOVATION...