RECOMMENDED SEPARATION DISTANCES
Recommended separation distances between portable and mobile RF communications
equipment and the Normatec 3 control unit.
Rated
maximum
output power
of transmitter W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance (d) in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
EQUIPMENT CLASSIFICATION
• Protection against electric shock: Class II/internally powered equipment
• Degree of protection against electric shock: Type BF applied part (control unit, leg,
arm, and hip attachments)
• Ingress protection: IP21
• Equipment not suitable for use in the presence of a flammable anesthetic mixture
with air or with oxygen or nitrous oxide
• Continuous operation
TROUBLESHOOTING
Problem
The system
does not start
The attachments
(leg, arm, or hip)
do not inflate
The system
stopped
pumping
Air leak
message: ERR
Low Battery
Separation distance according to frequency of transmitters in meters
150kHz – 80MHz
d=1.2√P
0.12
0.38
1.2
3.8
12
Possible Causes
Power is not turned
on Power supply is not
securely connected
Faulty electrical wall
outlet
The session has not
been started The
hose is not securely
connected The
attachments or hose
have been damaged
The hose is not
securely connected
The attachments have
been damaged
Air leak
Battery needs to be
charged
80MHz to 800MHz
d=0.35√P
0.12
0.38
1.2
3.8
12
Solutions
Press the power button to turn the control
unit on.
Check that the battery is charged or that
the charger is securely connected to the
Normatec 3 control unit and the electrical
outlet. Check that the wall outlet works.
Tap the start button to start the session.
Check that the hose is securely connected
to the Normatec 3 control unit and that the
attachments are securely connected to the
junction box. Check that the attachments
and/or hose do not have an air leak.
Check that the hose is securely connected
to the Normatec 3 control unit and that the
attachments are securely connected to the
junction box. Check that the attachments
do not have an air leak.
Check for leaks in the hose or attachment.
Check that the connectors are firmly
connected. if using only one attachment,
make sure blocking plug is fully seated in
junction box.
Plug in the control unit to charge the battery.
800MHz to 2.7GHz
d=0.7√P
0.23
0.73
2.3
7.3
23