Endosee
Hysteroscope - User Manual
®
11. After you copy files or folders to the hard drive, it
is useful to view the hard drive to confirm that the
files are copied.
12. Do not erase photos from your memory card
until you are positive you've saved them on the
hard drive!
8
Maintenance
8.1 Cleaning/Disinfecting Instructions
After each patient use, follow the combined cleaning/
disinfecting instructions provided below for the
Handheld Monitor. If the Handheld Monitor was
contaminated with blood or body fluids during use,
you must follow the intermediate level instructions. If
the Handheld Monitor was not contaminated with
blood or body fluids, follow the low level instructions.
8.2 Intermediate Level Combined
Cleaning/Disinfecting Instructions
•
After each patient use, power off and unplug all
electric connections.
•
Place Handheld Monitor on a clean, dry,
disinfected or sterile basin or surface.
WARNING:
provided for the level of contamination of
the device.
CAUTION:
Hysteroscope should be opened for
cleaning or disinfecting.
CAUTION:
disinfection process, orient the connector in a
manner that prevents liquid from entering the
electrical connection area of the connector
(e.g., with the opening to the electrical
connections facing upward). Do not wet the
opening or inside of connector housing the
electrical connection components.
Follow all cleaning procedures
No portion of the Endosee
Throughout the cleaning/
28