Installation and commissioning
Venting the safety storage cabinet > Connection to an exhaust air system
Fig. 17: Mounting the plinth panel
1
Three-part plinth panel
Mounting the plinth panel
Personal:
n
Technical specialist employees
1.
Connect the side pieces of the plinth panel to the front cover (A-C).
2.
Push the three-part plinth panel from the front underneath the cabinet (D-F).
3.
Connect the three-part plinth panel using the spring clip to the front adjustable feet.

6.6 Venting the safety storage cabinet

6.6.1 Connection to an exhaust air system
WARNING!
Insufficient safety storage cabinet air circulation
A lack of or insufficient air exchange can lead to formation of an explosive atmosphere in the safety storage
cabinet.
This may result in death or serious injury.
In a safety storage cabinet with ventilation system, a minimum hourly air exchange must take place that is
at least 10 times the internal volume of the cabinet when the doors are closed.
Installation of industrial ventilation and connection to an existing exhaust air system must be carried out by a
qualified company and is not a service provided by DÜPERTHAL.
CLASSIC line, COMPACT line 101