6.4
Cleaning the spray tube
• Switch off the unit and wait until the trimmer disc stops.
• Close the water shut-off valve of the water supply.
• Pull the model table upwards from the ball head at the front edge�
• Open the door locks (Fig� 8) and remove the trimmer door�
• Unscrew the spray tube (18) backwards on the curved side�
• Place the spray tube in descaling solution overnight�
• Reinsert the spray tube by pressing the spray tube into both holders�
6.5
Cleaning the water connection sieve
• Disconnect the power plug�
• Close the water shut-off valve of the water supply.
• Remove the water hose (9, Fig� 2) from the water connection (12, Fig� 2) of
the unit�
• Remove the sieve with a tongs and clean it�
• Place the sieve back into position�
6.6
Replacing the flange
• Switch off the unit and wait until the trimmer disc stops.
• Disconnect the power plug�
• Open the door locks (Fig� 8) and remove the trimmer door�
• Detach the trimmer disc (Chap� 5�4)�
• Screw the puller (11) into the flange (23) until the flange is detached from the
shaft�
Do not remove the flange by hand. Only the puller may be used!
• Remove the flange.
• Unscrew the puller again from the flange.
• If necessary, clean inside (behind the flange).
• If necessary, clean the hub�
• Attach the new flange >>> Note the engagement groove.
• Press the flange by hand onto the motor shaft.
• Attach the trimmer disc again (Chap� 5�4)�
• Attach the trimmer door and close the door locks�
• Connect the power plug�
6.7
Spare parts
You can find wear parts and spare parts in the spare parts list available online at
www�renfert�com/p918�
Enter the following item number: 18070000.
The components excluded from the warranty (wear parts, consumables) are
marked in the spare parts list�
The serial number, date of manufacturing and equipment version are shown on
the nameplate of the equipment�
- 9 -