Rectangular cuvette tests
NANOCOLOR
rectangular cuvette tests can be called up in the device in various ways. They
®
cannot be called up via a barcode. The measurement menu of the rectangular cuvette tests
can be called up by selecting the corresponding test from the list with rectangular cuvette
tests
or by entering the test number
requested. After pressing
insertion of the cuvette with the sample solution is requested. Confirming with
result in the measurement window. Various sample information can be entered using
the icons below the result field. Pulling out the cuvette or exiting the measurement menu via the
remaining icons in the task bar ends the measurement procedure and saves the result in the
measurement value memory of the device.
Scan
The spectrophotometer offers the option of recording the absorbance of a sample across
the wavelength range of 190 – 1100 nm (for NANOCOLOR
NANOCOLOR
VIS II). A scan with the halogen and/or the deuterium lamp is performed,
®
depending on the selected wavelength range and the respective device type. The measurement
is performed against a zero solution. After selecting the menu using
entered. The start and end wavelength of the measurement must be selected.
UV/VIS II
Scan: Blaue Lösung
420.0
463.0
506.0
549.0
592.0
635.0
1.229
1.104
0.979
0.855
0.730
0.605
0.480
0.355
0.321
0.106
-0.019
0010
Probenehmer Probeort
Kommentar
Fig. 13: Scan
Quickstart Guide NANOCOLOR
1
4
7
the measurement of the zero solution is performed and then the
Open Mode, 01.03.2016, 10:57
678.0
721.0
764.0
807.0
850.0 nm
1.229
1.104
0.979
0.855
0.730
0.605
0.480
0.355
0.321
0.106
-0.019
E
Null
Messen
. Insertion of the cuvette with the zero solution is
2
3
5
6
8
9
® UV
After confirming with
window appears and the insertion of the zero
solution is requested. Then the insertion of the
sample solution is requested. A graph to show
the absorbance over the wavelength opens.
The scan over the entire wavelength range can
be tracked live. Various sample information can
be entered using
result field. Pulling out the cuvette or exiting the
measurement menu via the remaining icons in
the task bar ends the measurement procedure
and saves the result in the measurement value
memory of the device. The menu for the scan
analysis can be accessed by calling up an
individual "scan" result in the measuring device
memory and then pressing the
Alternatively the scan analysis menu can be
reached by selecting the entry "scan analysis"
from the selection list of the
a scan.
/